Volume 27 Issue 3
Mar 2015
Turn off MathJax
Article Contents
CHEN Yi, HUANG Yong-Jun, FEI Teng. Calibration of Standard Hydrophones in the Frequency Range of 20 Hz to 200 kHz at 10 MPa Hydrostatic Pressure[J]. Chinese Journal of High Pressure Physics, 2013, 27(3): 454-460. doi: 10.11858/gywlxb.2013.03.021
Citation: CHEN Yi, HUANG Yong-Jun, FEI Teng. Calibration of Standard Hydrophones in the Frequency Range of 20 Hz to 200 kHz at 10 MPa Hydrostatic Pressure[J]. Chinese Journal of High Pressure Physics, 2013, 27(3): 454-460. doi: 10.11858/gywlxb.2013.03.021

Calibration of Standard Hydrophones in the Frequency Range of 20 Hz to 200 kHz at 10 MPa Hydrostatic Pressure

doi: 10.11858/gywlxb.2013.03.021
  • Received Date: 13 May 2011
  • Rev Recd Date: 28 Sep 2011
  • Issue Publish Date: 15 Jun 2013
  • This paper describes the coupler reciprocity method and free-field reciprocity method used for standard hydrophones calibration in the frequency range of 20 Hz to 200 kHz at 10 MPa hydrostatic pressure, and the coupler reciprocity calibration facility working in the frequency from 20 Hz to 3.15 kHz and free-field reciprocity calibration facility working in the frequency from 3.15 kHz to 200 kHz was established on these methods. Standard hydrophones RHC14 and RHS30 were calibrated separately using these facilities, and their sensitivities at 10 MPa hydrostatic pressure was presented. Analyze of measurement uncertainties proved that the expand uncertainty (at k=2) of coupler reciprocity calibration facility was 0.5 dB, and free-field reciprocity calibration facility was 0.7 dB (below 100 kHz) or 0.9 dB (100 kHz to 200 kHz).The construction of these facilities and capabilities sets the test foundation for deep-diving underwater sound devices.

     

  • loading
  • van Buren A L, Blue J E. Calibration of underwater acoustic transducers at NRL/USRD, power transducer for sonics and ultrasonics [C]//Proceedings of the International Workshop. Toulon, France, 1990: 221-241.
    Beamiss G A, Hayman G, Robinson S P. The provision of standards for underwater acoustics at simulated ocean conditions by use of the NPL acoustic pressure vessel, CMAM 76 [R]. London, UK: The National Physical Laboratory , 2001.
    International Electrotechnical Commission. IEC 60565: Underwater acoustics-hydrophones - calibration in the frequency range 0. 01 Hz to 1 MHz [Z]. Geneva, Switzerland: International Electrotechnical Commission, 2006.
    Chen Y, Chen W H, Yuan W J, et al. The development of free-field underwater acoustic pressure standard at high hydrostatic pressure [J]. ACTA Metrologica Sinica, 2008, 29(3): 257-261. (in Chinese)
    陈毅, 陈卫华, 袁文俊, 等. 高静水压下自由场水声声压计量标准装置研究 [J]. 计量学报, 2008, 29(3): 257-261.
    BIPM, IEC, IFCC, ISO, IUPAC, IUPAP, OIML. Guide to the expression of uncertainty in measurement [Z]. Geneva, Switzerland: International Organization for Standardization, 1995.